Pacemaker
227 490 02-C Marquette Responder® 3000  39
ECG
II
I
III
15.07.1998    09:05:00
160 /   40
bpm
manual
62
ECG
SpO2
etCO2
QRSPulse
Tone
OFF
Next
Menu
Pacer
Dem.
60
0 
mA
P/min
Figure 6-2. Pacer window
•
 
Select a low pacer current output with 
+ —
mA
,
e.g. 20 mA (19, Figure 6-1).
•
 
Select the required pacer rate with 
+ —
P/min
.
•
 
Now increase the pacer current output slowly
with 
+ —
mA
 until the heart reliably responds to
the stimulation.
•
 
Increase the pacer current output another 5 mA
to ensure continued stimulation.
•
 
Verify the stimulation by watching the ECG on
the display.
•
 
You can press 
Start
Pause
 to pause the stimulation and
press 
Start
Pause
 again to resume pacing with the same
settings.
•
 
After therapy, turn off the pacemaker before
removing the pace pads carefully.
(To turn off the pacemaker press the 
On
Off
 
button
for at least 2 seconds (this is to prevent that the
pacemaker is turned off inadvertently.) For this
reason the Marquette Responder® 3000 cannot
be turned off while the pacemaker is on.