Loading...
Pacemaker
227 490 02-C Marquette Responder® 3000 39
ECG
II
I
III
15.07.1998 09:05:00
160 / 40
bpm
manual
62
ECG
SpO2
etCO2
QRSPulse
Tone
OFF
Next
Menu
Pacer
Dem.
60
0
mA
P/min
Figure 6-2. Pacer window
Select a low pacer current output with
+
mA
,
e.g. 20 mA (19, Figure 6-1).
Select the required pacer rate with
+
P/min
.
Now increase the pacer current output slowly
with
+
mA
until the heart reliably responds to
the stimulation.
Increase the pacer current output another 5 mA
to ensure continued stimulation.
Verify the stimulation by watching the ECG on
the display.
You can press
Start
Pause
to pause the stimulation and
press
Start
Pause
again to resume pacing with the same
settings.
After therapy, turn off the pacemaker before
removing the pace pads carefully.
(To turn off the pacemaker press the
On
Off
button
for at least 2 seconds (this is to prevent that the
pacemaker is turned off inadvertently.) For this
reason the Marquette Responder® 3000 cannot
be turned off while the pacemaker is on.