EasyManua.ls Logo

GE Marquette Responder 3000

GE Marquette Responder 3000
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Test Discharge
72 Marquette Responder® 3000 227 490 02-C
3
2
Analyse
Sync
Figure 15-7. Buttons to deliver the shock
Caution
Equipment Damage — Do not trigger more than 3
consecutive test discharges (or internal safety
discharges) with the maximum energy of 360 J
with 15 minutes. Otherwise the device may reach
inadmissible high temperatures.
Note
If the shock is not delivered within 30 seconds of
charging, an internal safety discharge will be
initiated.
Now trigger the shock within 30 seconds. To do
so, simultaneously press the two buttons on the
paddles (or
and on the control panel,
when using adhesive pads) (Figure 15-7).
After the shock release, the audio signal stops and
the delivered energy is displayed for about 6
seconds. According to the requirements of the IEC
/ AAMI standards, this value must be between 306
and 414 Joules. At the same time, the defibrillator
prints a 14-second ECG strip (incl. history of 4
seconds), if this function is enabled.
If the discharge circuit is interrupted (electrodes
not properly inserted in their compartments, cable
defect), a safety discharge will be initiated 200 ms
after release of the test discharge.
Switch off the Marquette Responder® 3000
(energy selector to "Off").
Incorrect Shock Delivery, Error Messages
If the energy of the test discharge is not within the
specified limits, a defibrillation is possible all the
same (it is the user's decision whether or not to
employ the defibrillator). However, the device
must be immediately checked and repaired by a
service technician.

Related product manuals