8
8
Conguring for a Probe and Wedge
Access the probe and wedge settings by pressing to activate the
PROBE menu. Wedge features are adjusted as follows:
Wedge Offset Z – PROBE then WDGE GEO (user measured
from Probe Index Point to contact surface – set to 0 when no
wedge is installed)
Wedge Angle – PROBE then WDGE GEO ( user measured
wedge angle – set to 0 when no wedge is installed)
Wedge Front – PROBE then WDGE GEO (user measured
from Probe Index Point to wedge’s front edge – required to use
the origin offset feature)
Origin Offset – PROBE then OFFSET (user measured from
wedge’s front edge to the target)