PT878 Quick Start Guide 13
July 2003
Pipe Parameters (cont.)
Note: If you do not know the outside diameter, refer to Sound
Speeds and Pipe Size Data (914-004) or use the On-Line
Help function (refer to the Users Manual).
b. In either case, the next prompt asks for the wall thickness.
Type the value (in mm or in.) into the box, and press
[ENTER] to confirm your choice.
Note: If you have selected certain pipe materials (carbon or
stainless steel, cast iron, PVC and CPVC), you have the
option of entering the pipe dimensions by a standardized
schedule. Refer to Chapter 3 of the User’s Manual to
learn more about this option.
4. Press the [
S] key twice to return to the Pipe tab.
You have completed entering data in the Pipe tab. Proceed to the
Fluid tab.
Fluid Parameters
This section describes how to enter parameters for most common
fluids. To step through each parameter, press the [
T] key.
1. From the Pipe tab, press the [X] arrow key twice to reach the
Fluid tab, and press
[ENTER].
2. The first prompt asks you to select whether or not you want
Tracking Windows. This feature is used when the process
fluids soundspeed fluctuates considerably. (Default operation
is “No.”)
a. Use the [
W ] and [X] keys to scroll to the appropriate
radio button.
b. Press
[ENTER] to confirm your selection.