• Maximumrelativehumidity90%upto31°Cdecreasinglinearlyto50%at40°C
• Extremesoftemperaturemayrequirerecalibrationoftheunitforoptimum
performance
• Iftheinstrumenthasbeenstoredinacoldenvironmentthenitshouldbeallowed
to come to thermal equilibrium for 2 to 3 hours before operation so that start up
calibration is not compromised by condensation.
• Theinstrumentmustbeplacedonastable,levelbenchortablecapableoftaking
its weight with sufficient space around the instrument for ventilation to circulate
freely.
• Theequipmentmustbeconnectedtothelocalsupplyoutletusingtheprovided
power cables, compatible voltages are shown in the table below. Replace power
inlet fuses only with the same type and rating as follows
Instrument Type Power requirements Replacement Fuse
Ultrospec 10 units 100–240 V AC 50–60 Hz 3 A (UK plug only)
Novaspec III and Novaspec Plus 100–240 V AC 50–60 Hz 3 A (UK plug only)
GeneQuant100&1300units 100–240 V AC 50–60 Hz 3 A (UK plug only)
Ultrospec 2100 units 85–264 V AC, 50–60 Hz T 2 A H 250 V AC (Anti-Surge,
High breaking capacity)
Ultrospec 7000/8000/9000 units 85–264 V AC, 50–60 Hz T 1.6 A H 250 V AC (Anti-Surge,
High breaking capacity)
SimpliNano units 100–240 V AC 50–60 Hz 3 A (UK plug only)
NanoVue Plus units 90–250v, 50/60Hz, Max 30 VA 3 A (UK plug only)
• Theinstrumentshouldbepositionedsothatthepowercablemaybereadily
removed in the event of a hazard or malfunction occurring.
• Sitetheinstrumentinanatmospherefreefromdustandcorrosivefumes.
• Ultrospec2100instrumentshavetheon/offswitchonthebackoftheinstrument
on the right hand side. Ultrospec 7000/8000/9000 instruments have the on/off
switch on the left hand side of the instrument. Ultrospec 10, Novaspec III/Plus,
GeneQuant100/1300,SimpliNanoandNanoVueinstrumentshaveanon/off
button on the keypad – to disconnect power from these units remove the power
lead from the instrument. The instruments will automatically perform some start
up self diagnostic checks, please wait for these to finish before attempting to use
the equipment.
• Thespectrophotometermaybefittedwitharangeofaccessoriesforspecific
applications.FortheGeneQuant100/1300andNanoVuePlustheseincludeabuilt
in printer, bluetooth or SD card accessory. The SimpliNano has an optional built-
in printer. The Ultrospec 2100/7000/8000/9000 have a range of accessories that
can be fitted in the sample compartment including a sipper, long path length cell
holder and thermostatted cell holder. Full instructions on fitting accessories are
within the instrument or accessory user manual.