Product Description and Specifications
1–5
1.4 Specifications
Specifications are nominal and are subject to change without notice.
Factory settings
Setting Range Factory setting
High SpO
2
alarm limit 51 to 100% or OFF (— —) OFF
Low SpO
2
alarm limit 50 to 99% or OFF (— —) 85
High pulse rate alarm limit 30 to 235 bpm or OFF (— — —) 130
Low pulse rate alarm limit 30 to 235 bpm or OFF (— — —) 40
Backlight ON or OFF ON
Pulse beep volume 0 (OFF), 1, 2, 3, or 4 * 2
Alarm volume 1 (low), 2, 3, or 4 * 2
Line power filter 50 Hz or 60 Hz 60 Hz
* For more information, see Audio later in this section.
Measurement
General
Pulse oximetry sensors: OxyTip+ sensors only
Method: red and infrared light absorption
Red LED wavelength range: 650 to 670 nm
Infrared (IR) LED wavelength range: 930 to 950 nm
Average power: ≤ 1 mW
SpO
2
Calibrated for functional oxygen saturation
Calibration range: 70 to 100%
Measurement and display range: 1 to 100%
Display resolution: 1%
First reading, full accuracy: ≤ 10 seconds
Accuracy, A
rms
(root mean square of paired values; previously represented by ± 1 SD):
70 to 100% ± 2 digits (without motion)
70 to 100% ± 3 digits (during clinical motion)
1
70 to 100% ± 2 digits (during clinical low perfusion)
Below 70% unspecified
SpO
2
measurement accuracy is based on deep hypoxia studies using OxyTip+
sensors on volunteer subjects. Arterial blood samples were analyzed simultaneously
on multiple CO-oximeters.
NOTE: Accuracy may vary for some sensors; always check the instructions for the
sensor.
1
Applicability: OxyTip+ Adult/Pediatric and AllFit sensors.