IllustrationActionStep
Fit the two small lugs on the cover in the holes at the
front and rear of the cell holder.
7
Lower the cover over the cell holder.8
Installing UV Flow Cell ÄKTApilot
1/2/5 mm, and Industrial Flow
Cell 8 mm and 1"
ActionStep
Unpack the flow cell. Do not remove the black protective
caps on the optical fiber transmitter and receiver connector.
1
Choose the optical path length of the flow cell. The optical
path length can be changed with shim plates.
2
Attach the flow cell to an appropriate flow cell holder.3
Connect the inlet and outlet tubing to the flow cell.4
Connecting the optical fibers
WARNING
High intensity UV light. This product uses high intensity ultra-violet
light. Do not disconnect the optical fibers while the lamp is on.
Monitor UVis-920 Operating Instructions 29055049 AD 31
4 Installation
4.5 Installing the flow cells