Analog out connector
1
2
3
4
5
6
FunctionSignalPin
0 V to 1 V signal+1
Ground-2
Not used3-6
Connecting to recorder
Connect the recorder to the miniDIN-socket Analog out 0-1 V using the cable supplied.
The absorption signal is available on a channel using the following wires:
Wire 1: signal (+)
Wire 2: ground (-)
The signal cable is delivered with protective covers on each wire. Do not remove
the protective covers from unused connections as a short circuit may disturb
the measurements. Set the recorder to 0 V to 1 V input, full scale, 0 V offset.