Connecting to auxiliary
equipment (if used)
NOTICE
Any computer used with the equipment must comply with IEC
60950 and be installed and used according to the manufacturer's
instructions.
Connect any auxiliary equipment to the 9-pole D–Sub female Remote connector (5 V
TTL signals only).
1
2
3
4
5
6
7
8
9
FunctionSignalPin
Signal. Maximum load 50 mA5V (Out)1
Not used2
Not used3
Negative edge zeroes AU-valueAuto Zero (In)4
Common ground for all signalsGnd5
Goes low at error. Goes high again when OK is
pressed
Error (Out)6
0=UVis-920 is in Run-modeState (Out)7
1=UVis-920 is in End-mode
Monitor UVis-920 Operating Instructions 29055049 AD 35