DisplayActionStep
Use the arrow keys to set the value. Use the fixed
values between 5.12 seconds and 0.08 seconds.
2
Min. peak width at half
height (s)
Time constant (s)
(approximate)
Averaging time (s)
322.05.12
161.02.56
8.00.51.28
3.20.20.64
1.60.10.32
0.80.050.16
0.50.030.08
Setting cell serial number, cell
type and cell path length
It is possible to store cell data in the instrument for up to five different cells. The data is
uniquely defined by each cells serial number.
DisplayActionStep
Select menu Set Cell Serial Number, press OK.1
Enter the serial number of the cell (see the label on
the flow cell). Press OK. The serial number must be
1000 or higher.
2
Select menu Set Cell Type. Check the cell type value.
If the value is correct, use the arrow keys to go to
next menu. If the value has to be changed, press OK
to change it.
3
Select menu Set Cell Path Length, press OK.4