4°C to 40°CEnvironment
20%–95% relative humidity (noncondensing)
84–106 kPa (840-1060 mbar) atmospheric
pressure
< 70 dB ANoise emission
1
Measured with water at 1 ml/min, time constant 1 second, 10 mm flow cell.
2
Typical values at room temperature after 2 hours with lamp on.
Physical data
Xenon flash lampLight source
> 4000 hoursLamp lifetime
Stand alone or via Remote connectorControl
35 VAPower consumption
100–240 V AC ±10%, 50/60 HzPower requirement
0–1 V full scale, overrange functionAnalog output
5 V, 1 mA current sinking, lamp on/off, autozero, event
mark
Digital inputs
2 rows with 20 characters eachDisplay
200 x 160 x 262 mm without cell holderDimensions (H x W x D)
200 x 300 x 262 mm with cell holder
5.1 kgWeight
IP 20Degree of protection
< 70 dB ANoise emission
UV flow cell, 2 and 10 mm
100 ml/minRecommended maximum flow
rate
2 MPa (20 bar, 290 psi)Maximum pressure
Maximum 0.5 bar at 2 ml/min with water at 25°CBackpressure
62 Monitor UVis-920 Operating Instructions 29055049 AD
8 Reference information
8.1 Technical specifications