4°C to 40°CLiquid temperature range
2 mmOptical path length, 2 mm cell
10 mmOptical path length, 10 mm cell
2 μLCell volume, 2 mm cell
8 μLCell volume, 10 mm cell
PTFE (polytetrafluoroethylene)Wetted materials
PEEK (polyetheretherketone)
Titanium (palladium alloy)
Quartz (synthetic fused silica)
1–13, 13–14 (< 1 days exposure)pH stability range
The wetted parts are resistant to organic solvents
and salt buffers commonly used in chromatogra-
phy of biomolecules, except 100% ethyl acetate,
100% hexane, and 100% tetrahydrofuran (THF)
or 15% THF in acetonitrile
Chemical resistance
UNF 10-32 fingertight connectors for tubing with
1/16" outer diameter
Tubing connections
UV Flow Cell ÄKTApilot 1/2/5 mm
800 ml/minRecommended maximum flow
rate
2 MPa (20 bar, 290 psi)Maximum pressure
Max. 0.05 bar at 400 ml/min with water at 25°CBackpressure
4°C to 40°CLiquid temperature range
1, 2 and 5 mmOptical path length
300 μLCell volume
PEEK (polyetheretherketone)Wetted materials
Titanium (palladium alloy)
FFKM (perflorelastoner) or EPDM
Quartz (synthetic fused silica)
1–13, 13–14 (< 1 days exposure)pH stability range
Monitor UVis-920 Operating Instructions 29055049 AD 63
8 Reference information
8.1 Technical specifications