The wetted parts are resistant to organic solvents
and salt buffers commonly used in chromatogra-
phy of biomolecules, except 100% ethyl acetate,
100% hexane, and 100% tetrahydrofuran (THF)
Chemical resistance
5/16"Tubing connections
Industrial flow cell 8 mm
*Recommended maximum flow
rate
2 MPa (20 bar, 290 psi up to 40°CMaximum pressure
Max. 0.1 bar at 1800 l/h with water at 25°CBackpressure
4°C to 80°C (at max 0.1 MPa)Liquid temperature range
1, 2 and 5 mmOptical path length
2.5 mlCell volume
PEEK (polyetheretherketone)Wetted materials
Titanium (palladium alloy)
Quartz (synthetic fused silica)
PFR-91 or EPDM
1–13, 13–14 (< 1 days exposure)pH stability range
The wetted parts are resistant to organic solvents
and salt buffers commonly used in chromatogra-
phy of biomolecules, except 100% ethyl acetate,
100% hexane, and 100% tetrahydrofuran (THF) or
15% THF in acetonitrile
Chemical resistance
TC25 connectorTubing connections
Industrial flow cell 1"
*Recommended maximum flow
rate
1 MPa (10 bar, 145 psi) up to 40°CMaximum pressure
Max. 0.4 bar at 600 l/h with water at 25°C.Backpressure
64 Monitor UVis-920 Operating Instructions 29055049 AD
8 Reference information
8.1 Technical specifications

Related product manuals