Vscan Air Maintenance
6-12 Vscan Air – User Manual
Direction GP092020-1EN
Rev 18
Low-level/Intermediate level disinfection with wipes
1. Use a clean set of gloves.
2. Hold the Vscan Air CL with the large lens facing away from
your body.
3. Wipe the Vscan Air CL from the large lens to the small lens,
slightly rotating the Vscan Air CL after each wiping pass.
4. Turn the Vscan Air CL so that the small lens is facing away
from your body.
5. Use a new wipe and apply to the small lens. Now, wipe the
Vscan Air CL from the small lens to the large lens, slightly
rotating the Vscan Air CL after each wiping pass.
6. Wrap a clean wipe around a soft nylon bristle brush to
access crevasses on the surface of the Vscan Air CL. Do
not use the brush on the Vscan Air CL lenses.
7. Once the Vscan Air CL has been completely wiped, use
additional wipes and continue wiping the Vscan Air CL as
needed to ensure all surfaces remain wet for the required
exposure time listed on the disinfectant manufacturer’s
label.
8. Dry all surfaces of the Vscan Air CL using a sterile, lint-free,
soft wipe or cloth. Blot or pat the lenses dry.
9. After each use, inspect the lenses, and housing of the
Vscan Air CL. If the Vscan Air CL is damaged, do not place
it into any liquid (e.g. for disinfection) and do not use it until it
has been inspected and repaired/replaced by a GEHC
Service Representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vscan Air and is the answer not in the manual?

GE Vscan Air Specifications

General IconGeneral
Battery LifeUp to 50 minutes of scanning
TypeHandheld Ultrasound
DisplayVia mobile device (not included)
ConnectivityWi-Fi
Probe TypeDual-probe
Charging TimeApproximately 2 hours
ChargingUSB-C
Water ResistanceIP67
Transducer TypePhased Array and Linear Array
Imaging ModesColor Doppler
Operating SystemAndroid, iOS
StorageVia mobile device (not included)
CompatibilityCompatible with Android and iOS devices

Related product manuals