EasyManuals Logo

GE Vscan Air User Manual

GE Vscan Air
407 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #366 background imageLoading...
Page #366 background image
Vscan Air Maintenance
6-42 Vscan Air – User Manual
Direction GP092020-1EN
Rev 18
Transducer Element test
The transducer element test transmits/receives on each element
to determine if the element working as intended. If the Element
test fails, the failing element(s) are reported in the system logfile.
The transducer element test is part of the probe diagnostics test
which is found via the app Menu -> Diagnostics -> Run
Diagnostics.
The element test is then executed as part of a predefined test
list of diagnostic tests.
NOTE: If diagnostics fails, GEHC service will have to look through the
logfile to determine which test failed. If the transducer element
test failed, service will then be able to identify the failing
transducer element(s) and guide on further actions needed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vscan Air and is the answer not in the manual?

GE Vscan Air Specifications

General IconGeneral
Battery LifeUp to 50 minutes of scanning
TypeHandheld Ultrasound
DisplayVia mobile device (not included)
ConnectivityWi-Fi
Probe TypeDual-probe
Charging TimeApproximately 2 hours
ChargingUSB-C
Water ResistanceIP67
Transducer TypePhased Array and Linear Array
Imaging ModesColor Doppler
Operating SystemAndroid, iOS
StorageVia mobile device (not included)
CompatibilityCompatible with Android and iOS devices