Safety
2-14 Vscan Air – User Manual
Direction GP092020-1EN
Rev 18
Electromagnetic immunity
Table 2-2: Electromagnetic immunity (Part 1)
Guidance and manufacturer’s declaration – electromagnetic immunity.
The Vscan Air CL probe is intended for use in the electromagnetic environment below.
The customer or the user of the Vscan Air CL should assure that it is used in such an
environment.
Immunity test EN/IEC 60601
test level
Compliance
level
Electromagnetic
environment –
guidance
Electrostatic
discharge (ESD)
EN/IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least
30%.
Electrical fast
transient/ burst
EN/IEC 61000-4-4
±2 kV for
power-supply lines
±1 kV for input/
output lines
±2 kV for
power-supply
lines
±1 kV for input/
output lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
EN/IEC 61000-4-5
±1 kV line(s) to
line(s)
±1 kV line(s) to
line(s)
Mains power quality
should be that of a
typical commercial or
hospital environment.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vscan Air and is the answer not in the manual?

GE Vscan Air Specifications

General IconGeneral
Battery LifeUp to 50 minutes of scanning
TypeHandheld Ultrasound
DisplayVia mobile device (not included)
ConnectivityWi-Fi
Probe TypeDual-probe
Charging TimeApproximately 2 hours
ChargingUSB-C
Water ResistanceIP67
Transducer TypePhased Array and Linear Array
Imaging ModesColor Doppler
Operating SystemAndroid, iOS
StorageVia mobile device (not included)
CompatibilityCompatible with Android and iOS devices

Related product manuals