Cleaning and disinfection
Vscan Air – User Manual 6-19
Direction GP092020-1EN Rev 18
Disclaimer: * Disinfection with Tristel Trio Wipes System may
cause discoloration of the lightbar, the button and the soft plastic
surrounding the battery door. Such color change does not
impact the performance, safety or lifetime of the probe. If color
change occurs due to disinfection with Tristel Trio Wipes
System, Vscan Air probes will not be repaired or replaced under
warranty or contract agreements.
Disclaimer: * Disinfection in trophon chamber may cause
discoloration of the lightbar, the button and the soft plastic
surrounding the battery door. Such color change does not
impact the performance, safety or lifetime of the probe. If color
change occurs due to disinfection in trophon chamber, Vscan Air
probes will not be repaired or replaced under warranty or
contract agreements.
Tristel Rinse Wipes Tristel Solutions Limited
Tristel Trio Wipes System * Tristel Solutions Limited
mikrozid sensitive or mikrozid alcohol free (liquid
and wipes)
Schulke & Mayr GmbH
mikrozid universal liquid and mikrozid universal
wipes
Schulke & Mayr GmbH
Protex ULTRA Disinfectant Wipes Parker Laboratories Inc
Reynard Premier Detergent & Disinfectant Wipes Reynard Health Supplies
Anioxyde 1000 Laboratoires Anios
Wip'Anios Excel Laboratoires Anios
Table 6-2: Chemicals compatible with the Vscan Air CL probe
Trade Name Manufacturer
Table 6-3: Products compatible with the Vscan Air CL probe
Trade Name Manufacturer
trophon2 * Nanosonics
trophon EPR * Nanosonics