Vscan Air Maintenance
6-26 Vscan Air – User Manual
Direction GP092020-1EN
Rev 18
Troubleshooting in Vscan Air CL
Table 6-5: Troubleshooting
Problem Possible cause Solution
Vscan Air CL has no power. When battery is discharged. Charge the Vscan Air CL probe
for at least 10 minutes and then
power on.
Battery defect or end of life. . Contact GEHC Service (see
‘Contact Information’ on
page 1-12)
Vscan Air CL is not charging. Defective battery or probe
hardware issue
Contact GEHC Service (see
‘Contact Information’ on
page 1-12)
Defective AC adapter. Contact GEHC Service (see
‘Contact Information’ on
page 1-12)
Defective wireless charger pad. Contact GEHC Service (see
‘Contact Information’ on
page 1-12)
Defective USB cable. Contact GEHC Service (see
‘Contact Information’ on
page 1-12)
Mains power is down. Ensure that the charger is
connected to a powered wall
outlet.
Temperature is outside the
specified limits.
Ensure the ambient temperature
is within the specified limits (see
‘Contact Information’ on
page 1-12)
Display screen is blank when the
device is powered on.
Connection broken during
software loading.
Contact GEHC Service (see
‘Contact Information’ on
page 1-12)
Parts of the image is missing
when scanning.
Channels are missing. Contact GEHC Service (see
‘Contact Information’ on
page 1-12)
No image displayed when
scanning.
Defective probe. Contact GEHC Service (see
‘Contact Information’ on
page 1-12)
Scan screen is not displayed. Battery may not have sufficient
charge.
Charge the Vscan Air CL probe
for at least 60 minutes.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vscan Air and is the answer not in the manual?

GE Vscan Air Specifications

General IconGeneral
Battery LifeUp to 50 minutes of scanning
TypeHandheld Ultrasound
DisplayVia mobile device (not included)
ConnectivityWi-Fi
Probe TypeDual-probe
Charging TimeApproximately 2 hours
ChargingUSB-C
Water ResistanceIP67
Transducer TypePhased Array and Linear Array
Imaging ModesColor Doppler
Operating SystemAndroid, iOS
StorageVia mobile device (not included)
CompatibilityCompatible with Android and iOS devices

Related product manuals