EasyManua.ls Logo

GE Vscan Air - Page 46

GE Vscan Air
407 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Safety
2-10 Vscan AirUser Manual
Direction GP092020-1EN
Rev 18
Electromagnetic Compatibility (EMC) (continued)
Any electrical device can unintentionally emit electromagnetic
waves. However, minimum device separation distances cannot
be calculated for such unspecified electromagnetic radiation.
When the ultrasound unit is used adjacent to or in close
proximity to other equipment the user should be attentive to
unexpected device behavior which may be caused by such
electromagnetic radiation.
The ultrasound unit is intended for use in the electromagnetic
environment specified in the tables below ‘Electromagnetic
emissions’ on page 2-13.
The user of the ultrasound unit should assure that the device is
used in such an environment.
NOTE: It is advised to keep a 30 cm separation distance in between the
Vscan Air CL probe and the display device running the Vscan
Air app.
WARNING
The use of accessories and cables other than those specified,
may result in increased electromagnetic emissions or
decreased electromagnetic immunity of the Vscan Air CL
probe.
WARNING
The Vscan Air CL probe should not be used adjacent or very
close to other equipment. The Vscan Air should be observed to
verify normal device behavior in the configuration in which it will
be used.
WARNING
Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the Vscan
Air device. Otherwise, it may degrade the performance of this
equipment.

Table of Contents

Related product manuals