Safety
2-16 Vscan Air – User Manual
Direction GP092020-1EN
Rev 18
Table 2-3: Electromagnetic immunity (Part 2)
Guidance and manufacturer’s declaration – electromagnetic immunity – for all medical
electrical equipment and medical electrical systems that not life-supporting
The Vscan Air CL probe is intended for use in the electromagnetic environment below.
The customer or the user of the Vscan Air CL should assure that it is used in such an
environment.
Immunity test IEC 60601 test level Compliance level
Conducted RF
IEC 61000-4-6
3Vrms
150 kHz to 80 MHz
3Vrms
Radiated RF and Proximity fields
from RF wireless
communications equipment
IEC 61000-4-3
10 V/m; 80 MHz to 2.7 GHz 80%
AM at 1 kHz
10 V/m; 80 MHz to 2.7 GHz 80%
AM at 1 kHz
385 MHz (18 Hz Pulse
Modulation)
27 V/m
450 MHz (FM +/ -5 kHz
deviation1 kHz sine or 18 Hz
Pulse Modulation)
28 V/m
710 MHz (217 Hz PM) 9 V/m
745 MHz (217 Hz PM) 9 V/m
780 MHz (217 Hz PM) 9 V/m
810 MHz (18 Hz PM) 28 V/m
870 MHz (18 Hz PM) 28 V/m
930 MHz (18 Hz PM) 28 V/m
1720 MHz (217 Hz PM) 28 V/m
1845 MHz (217 Hz PM) 28 V/m
1970 MHz (217 Hz PM) 28 V/m
2450 MHz (217 Hz PM) 28 V/m
5240 MHz (217 Hz PM) 9 V/m
5500 MHz (217 Hz PM) 9 V/m
5785 MHz (217 Hz PM) 9 V/m
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.