ActionStep
Install two ratchet clamps over the section of tubing between the bag as-
sembly probe port (welded into the disposable bag) and the female aseptic
connection device (ACD) connector attached to the probe port of the dispos-
able bag, at the location shown in the following illustration.
1
Using your fingers, loosely tighten the ratchet clamps until the teeth on the
clamp are engaged enough to hold the clamp in position (2 to 3 teeth en-
gaged).
2
NOTICE
Do not tighten the clamps too much, otherwise they
prevent the insertion of the probe into the disposable
bag.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD 169
6 Preparation
6.4 Prepare and install the probes
6.4.4 Insert probes into the disposable bag

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Xcellerex XDM 200 and is the answer not in the manual?

GE Xcellerex XDM 200 Specifications

General IconGeneral
ModelXcellerex XDM 200
Capacity/Nominal Volume200 L
Volume Range50-200 L
Mixing SpeedVariable
Vessel MaterialStainless steel
Single-Use LinerYes
Liner MaterialPolyethylene
MaterialStainless steel and polyethylene
SterilizationGamma irradiation for single-use components
Control SystemIntegrated control system
Power SupplyElectrical
TypeSingle-Use Mixer

Related product manuals