DenOptix QST
032-0265-EN Rev 12-2
Do not place DenOptix QST near:
X-ray equipment that is constantly energized (e.g., devices such as
fluoroscopes which are energized when not in use)
Magnetic resonance imaging systems
Large motor/electric generators
There is potential for electromagnetic inter
fer
ence between devices when using
electronic equipment. International standards exist to minimize this potential.
DenOptix QST has been tested against and compl
ies with the international
standard EN 60601-1-2, Medical Electrical Equipment - General Requirements for
Safety - Collateral Standard; Electromagnetic Compatibility Requirements and
Tests, class A.
Use only grounded electrical connections.
Connect the DenOptix QST scanner to a grounded electrical outlet between 100-240
volts AC.
Do not reuse the barrier envelopes.
Once used within the oral cavity, the barrier envelopes ar
e bio-contaminated and
should NEVER be reused. Please dispose of in accordance with page 5-3.
Imaging plates are toxic!
Never place an imaging plate in a patient’s
mout
h without enclosing it first in a
completely sealed barrier envelope. If a patient swallows an imaging plate, contact
a physician immediately. The physician must remove the imaging plate. Do not use
cracked, bent or chipped imaging plates.
Never open or override the lid lock when the sc
anner i
s in use or plugged in.
The scanner lid will lock during operation. In c
ase of power failure, unplug the
scanner and open the lid by depressing the lock mechanism as indicated on page
5-2.
Never use imaging plates with patients that might chew or swallow them.
If the patient bites or chews the plate and dam
ag
es the protective barrier
envelope, rinse the patient’s mouth with a large amount of water. If the patient
manages to swallow any of the blue surface of the plate, contact a doctor
immediately.
The DenOptix QST scanner is a Class 1 laser device.
CAUTION
Use of controls or adjustments or performance of procedures other than those
specified herein may result in hazardous radiation exposure. A fail-safe switch in
the carousel well and cover keeps the laser inactive as long as the cover is open.
Only a trained technician from a qualified Gendex dealer should remove the
cover from the scanner. Direct eye contact with the output beam from the laser
may cause serious damage and possible blindness.

Table of Contents

Other manuals for Gendex DenOptix QST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gendex DenOptix QST and is the answer not in the manual?

Gendex DenOptix QST Specifications

General IconGeneral
BrandGendex
ModelDenOptix QST
CategoryDental equipment
LanguageEnglish

Related product manuals