System Operating Instructions
032-0265-EN Rev 1 4-5
6. Wearing gloves, remove the
imaging plate in the barrier
envelope from the pati
ent’s
mouth. Wipe offany excess
saliva with a paper towel. After
each exposure, open the
barrier envelope at the “V” cut
in the center of the packet, with
a steady pull, and let the plate
fall into the transfer container
(Dark Box provided with the
system).
At the end of the intraoral
exam
ination, r
emove and
dispose of gloves and move to
a semi-darkened room or area. Remove lid and begin the process of loading the
imaging plates.
Do not touch the transfer container with
contaminated gloves.
7.
Remove your gloves and
wash any powder from your
hands. Mount the I/O
imagi
ng plates onto the
carousel by sliding the right
side of the imaging plate into
the correct size I/O mounting
hole in the carousel. Ensure
that the blue side faces out.
Repeat until all I/O imaging
plates are mounted.
8.
Open the carousel well
cover on the DenOptix QST
scanner a
nd insert the
carousel. Close the lid. You
are now ready to scan your
I/O imaging plates.