DenOptix QST
032-0265-EN Rev 1D-4
Guidance and manufacturer’s declaration – electromagnetic immunity
The Gendex QST is intended for use in the electromagnetic environment specified below. The
customer or the user of the Gendex QST should ensure that it is used in such an environment.
Immunity Test IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 150
kHz to 80
MHz
3 V/m 80
MHz to 2.5
GHz
3 V
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the Gendex QST, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
80 MHz to 800 MHz
800 MHz to 2,5 GHz