POCKIT Central Nucleic Acid Analyzer
21
Section 7. Cleaning and Shipping POCKIT Central
7.1 Cleaning and Maintenance
In case of spills and leakage of liquid, cleaning mode can be accessed from Cleaning /
Shipping Setup screen.
Note: Do not disassemble POCKIT Central for cleaning. This will void the warranty.
Go to Home > Cleaning / Shipping Setup
Select Cleaning / Shipping Setup on Home screen, and
press Clean the interior button.
Select OK to open the door.
Note:
1. Make sure nothing blocks the door.
2. Skip this step if the Door Opening/Closing Alert
alert is off.
Use dry or damp lint-free cloth/wipes to clean the tray
(circled area). Do not leave any solutions on the tray as
the electronic parts may be damaged. Wipe off any
visible stain on the sensors immediately.
Note: Not regularly cleaning the instrument may
decrease instrument reliability.