WindSonic Doc No 1405 PS 0019 Issue 28 November 2019
53
Kxx to Kxxxx - Minimum Direction Velocity
Item
Options
Command
Minimum Direction
Velocity
0.05m/s to 5m/s
K50
KXX to KXXXX
This feature allows the minimum wind velocity to be set at which display of the
direction is enabled in thousandth of a metre per second. The default is 50, which
means that any wind magnitude less than 0.05m/s will have the wind direction omitted
in Comma Separated Variable mode.
E.g. Q,,000.04,M,3C.
Or last wind reading at 0.05m/s or above repeated in Fixed Field Mode.
E.g. Q,237,000.04,M,3C.
Lx - Message Terminator
Item
Options
Command
Message terminator
<CR> <LF>
<CR>
L1
L2
Mx to Mxx - Message Format
Item
Options
Command
Message
Format
Gill, UV, Continuous
Gill, Polar, Continuous
Gill, UV, Polled
Gill, Polar, Polled
NMEA, IIMWV (NMEA status code)
Tunnel Continuous
NMEA, WIMWV (NMEA status code)
NMEA, WIMWV (Gill status code)
NMEA, IIMWV (Gill status code)
M1
M2
M3
M4
M5
M12
M14
M15
M16
Nx - Node Address
Item
Options
Command
Node Address
Q… -(A to Z)
N<Q>
If used in polled mode then it is recommended that ID’s A to F and K, M, N and P are not
used, as these characters can be present in the data string.