WindSonic Doc No 1405 PS 0019 Issue 28 November 2019
54
Ox - ASCII Output Format
Item
Options
Command
ASCII output format
Comma separated format (CSV)
Fixed field
O1
O2
Example data string for CSV data changing to low wind less than 0.05m/s (set for K50) and
error status code condition.
Q,156,000.05,M,00, 19
Q,,000.04,M,00, 2A (low wind, no direction figure output at 0.04m/s or lower).
Q,100,000.09,M,00, 16
Q,,,M,04, 24 (error data, no figures output in wind fields).
Example data string for Fixed Field data changing to low wind less than 0.05m/s (set for
K50) and error status code condition.
Q,160,000.05,M,00, 1C
Q,160,000.04,M,00, 1D (low wind, last direction figure at 0.05m/s or above retained).
Q,170,000.06,M,00, 1E
Q,999,999.99,M,04, 0A (error data, wind fields padded with 9’s).
Px to Pxx - Output Rate
Item
Options
Command
Output rate
1 per second
2 per second
4 per second
1 reading every 4 seconds
1 reading every 2 seconds
P1
P2
P3
P20
P21
Q - Measurement Mode
See Section 10.5.