Bravo pH Monitoring System 14 User Guide
chapter.fm 3/21/16 4:05 pm
5.5 x 7.75 inches (140 mm x 197 mm)
tube 1: pH 7.01 buffer solution
tube 2: sterile water
tube 3: pH 1.07 buffer solution
tube 4: sterile water
5. Check the expiration date on the capsule (next to on the label).
6. Without bending or kinking the delivery device, carefully remove the Bravo pH
capsule with delivery device from the external shipping box and then from the inner
pouch (Figure 4).
Figure 4. Remove Bravo pH capsule with delivery device from pouch.
7. Remove the capsule’s plastic cover, the reference sensor cover (soaker bulb cover),
and the magnetic clip (Figure 5). Set the magnetic clip aside.
!
Caution
Avoid bending or kinking the delivery device. Sharp bending or kinking
can damage the delivery device, which will require the device to be
discarded before use.
֠
Note
Keep the magnetic clip at least 2 meters (6 feet) away from the delivery
device so that it will not interfere with the capsule. If a procedure is
delayed, you can replace the magnetic clip on the capsule to return it
to an inactive state until needed. Do not discard the magnetic clip until
after the procedure has been performed.