Bravo pH Monitoring System 15 User Guide
chapter.fm 3/21/16 4:05 pm
5.5 x 7.75 inches (140 mm x 197 mm)
Figure 5. Remove Bravo pH capsule shipping components.
.
8. Check the soaker bulb for fluid and set aside.
֠
Note
The capsule automatically turns on when the magnetic clip is removed.
In addition, if the recorder is turned on, you can see the capsule status
LED blinking on the recorder.
֠
Note
Even if the soaker bulb doesn’t show the presence of liquid, the
capsule can calibrate as usual. Perform the capsule calibration as per
instructions. If the capsule fails to calibrate, contact your customer
support representative.
Open the plastic cover.
Remove magnetic clip.
Remove soaker bulb.