Osicus User Guide V8 09/09/2021 Page 41 of 46
Appendix 1: Guidance on the Electromagnetic environment in which to operate
the instrument (cont)
Guidance and manufacturer’s declaration – electromagnetic emissions
The Osicus is intended for use in the electromagnetic environment specified below. The customer or
the
user of the Osicus should assure that it is used in such an environment.
Electromagnetic environment - guidance
The equipment is suitable for use in a
professional healthcare facility environment only.
Harmonic emissions
IEC61000-3-2
Voltage fluctuations / flicker
emissions
IEC61000-3-3
Guidance and manufacturer’s declaration – electromagnetic immunity
The Osicus is intended for use in the electromagnetic environment specified below. The customer or the
user of the
Osicus should assure that it is used in such an environment.
Electromagnetic
environment -
guidance
Electrostatic discharge
(ESD)
IEC61000-4-2
± 2, 4, 6, 8 kV contact
± 2, 4, 8, 15 kV air
± 2, 4, 6, 8 kV contact
± 2, 4, 8, 15 kV air
Floors should be wood,
concrete
or ceramic tile. If
floors are
covered with
synthetic material,
the
relative humidity should be
at
least 30%.
Electrical fast transient /
burst
IEC61000-4-4
± 2 kV for power
supply lines
± 1 kV for input /
output lines
Mains power quality
should be
that of a
typical commercial or
hospital environment.
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality
should be
that of a
typical commercial or
hospital environment.