Osicus User Guide V8 09/09/2021 Page 45 of 46
7.2.17 Environmental conditions for transport and storage with no additional special measures
Temperature: -40 °C to +60 °C
Humidity: 30 to 90% RH
Pressure 98 kPa to 104 kPa
7.9.1 The Osicus Audiometer can be used by any medically trained technician, nurse or doctor,
who has either read this manual or who has been trained by a competent authority in its use.
7.9.2.1 The Osicus Audiometer health care professional or competent person in medical
environment, i.e. hospital, clinic.
7.9.2.3 The Osicus power is intended to receive power via a USB connection with a ISO 60950
compliant laptop/PC. The combination is considered to be an ME SYSTEM.
7.9.2.5 The Headset and Response Button are considered to be the applied parts.
7.9.2.7 The equipment should be positioned to enable it to be disconnected from the mains
supply quickly and easily.
7.9.2.10 Error Messages --- see Troubleshooting section in section
7.9.2.11 The Osicus software can be closed by clicking EXIT. The Osicus hardware can be switched
off by either of the following:
Removing the USB cable from the back of the Osicus
By disconnecting it from the PC USB socket
Powering down the PC
7.9.2.13 There are no parts which need routine maintenance or servicing other than general
“housekeeping”, such as keeping the equipment clean and cables free of twisting
7.9.2.15 The EXPECTED SERVICE LIFE of the Osicus has been evaluated and determined to be 10
years from the date of manufacture. After this time, the Osicus and the accessories, Headset,
Response Button and supplied PSU, can be returned to GM Instruments for dismantling and
disposal, while your PC/Printer should be returned to your supplier for dismantling and disposal.
These parts are subject to the WEEE directive, and should not be disposed of in landfill.
7.9.3.2 The only parts which are interchangeable by service personnel are the power module and
the mains cable. If these are changed, a Medical Electrical Safety test to the current standard
should be made prior to allowing release for use.
7.9.3.3 Circuit diagrams, component lists and parts lists are available on request, along with
email/telephone advice to service personnel trained and qualified to work on ME devices.
Modification of the Osicus Audiometer is not allowed.
7.9.3.4 If access to the Osicus circuit board is required:-
Disconnect the Osicus by disconnecting the USB cable linking it to the computer
Turn the unit over and remove the 4 feet on the base.
Turn the unit back over and remove the top panel.