Manufacturer’s declaration-electromagnetic immunity
The LS300 is intended for use in the electromagnetic environment (for home and professional
healthcare) specified below.
The customer or the user of the LS300 should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance level
Electromagnetic
environment-guidance
(for home and professional
healthcare environment)
Electrostatic
discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8 kV,±15
kV
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30%
Electrical fast
transient/burst
IEC 61000-4-4
+ 2kV for power supply
lines
+ 1kV for input/output
lines
+ 2kV for power supply
lines
Not applicable
Mains power quality should
be that of a typical home
healthcare environment.
Surge
IEC 61000-4-5
+ 0.5kV, +1kV line(s) to
line(s)
+ 0.5kV, +1kV,+ 2kV
line(s) to earth
+ 0.5kV, +1kV line(s) to
line(s)
Not applicable
Mains power quality should
be that of a typical home
healthcare environment.
Voltage Dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Voltage dips:
0 % U
T; 0,5 cycle
0 % U
T; 1 cycle
70 % U
T; 25/30 cycles
Voltage interruptions:
0 % U
T; 250/300 cycle
Voltage dips:
0 % U
T; 0,5 cycle
0 % U
T; 1 cycle
70 % U
T; 30 cycles
Voltage interruptions:
0 % U
T; 300 cycle
Mains power quality should
be that of a typical home
healthcare environment. If
the user of the LS300
requires continued operation
during power mains
interruptions, it is
recommended that the
LS300 be powered from an
uninterruptible power supply .
Power frequency(50, 60
Hz) magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
60 Hz
The LS300 power frequency
magnetic fields should be at
levels characteristic of a
typical location in a typical
home healthcare
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
*DuringDIPinterference,thepumpwilloutage.Theaircellsconnectedwithpumpstill
haveairinsidewhichwon’taffecttheuseandfunctionofthesystem.
*DuringDIP,pumpwillshowabnormalbutwon’taffectessentialperformanceandno
needtoworrythebasicsafety.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Graham Field LUMEX SELECT Series and is the answer not in the manual?

Graham Field LUMEX SELECT Series Specifications

General IconGeneral
BrandGraham Field
ModelLUMEX SELECT Series
CategoryIndoor Furnishing
LanguageEnglish

Related product manuals