Manufacturer’s declaration-electromagnetic immunity
The LS300 is intended for use in the electromagnetic environment (for home and professional
healthcare) specified below.
The customer or the user of the LS300 should assure that it is used in such and environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
(for home and professional healthcare
environment)
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Portable and mobile RF
communications equipment should be
used no closer to any part of the LS300
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.