After filling the bath adjust the water flow rate to the minimum which maintains a constant
water level when the water is boiling. Check that any relevant local water supply
regulations are complied with if connecting to a mains water supply.
5.4.5 Resetting the over temperature cut-out devices
Two fixed over temperature cut-outs prevent the heater from overheating in the case of a
low water level. The reset buttons for the over temperature cut-outs are on the side panel
and can be reset by removing the protective black cap and pressing the red button in.
The over temperature cut-outs may activate during normal operation. If this is the case
then they can be reset and the bath can continue to be used without compromising the
safety of persons or the surroundings. However, if the cut-outs are persistently activated
then it is recommended to have the bath checked by a competent person as soon as
possible.
6.0 Technical specifications
Operating conditions
Maximum relative humidity
80% R.H. in room temperatures up to 31°C
decreasing linearly to 50 % R.H. at 40°C
Electrical details
Mains supply: 230V @ 50/60 Hz or 120V @ 50/60 Hz
Polution degree: 2
Installation Category: II
Note: Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply
voltage