2. Loose laser control network connections.
3. Faulty power supply.
4. Faulty Interlock module.
5. Faulty Laser Control Card.
7.6.10 Alarm 4n10 : Power Supply Failed To Simmer Timeout
Meaning:
The power supply failed to simmer within the timeout period during the laser start up
sequence.
Possible causes:
1. High coolant water conductivity.
2. Broken lamp.
3. Faulty power supply.
7.6.11 Alarm 4n11 : Power Supply Simmer Under Power Fault.
Meaning:
The simmer current was above operating limits when the power supply was on.
Possible causes:
1. Broken lamp.
2. Incorrect lamp type.
3. Faulty power supply.
7.6.12 Alarm 4n12 : Power Supply Simmer Over Power Fault
Meaning:
The power supply simmer circuit is on when it should have turned off.
Possible causes:
1. Faulty Power Supply.
7.6.13 Alarm 4n13 : Power Supply Failed To Enable
Meaning:
The Interlock Module timed out waiting for power supply n to enter, and report that it had
entered the enabled state during the laser start up sequence.
Possible causes:
1. Incorrect laser type setting.
2. Loose laser control network connections.
3. Faulty power supply.
4. Faulty Interlock module.
5. Faulty Laser Control Card.
7.6.14 Alarm 4n14 : Power Supply Limit Fault
Meaning:
The output of the power supply is being driven to its limit.