LCR-800 Series User Manual
24
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
L 1.2345H
Q 0.6789
TESTING
F : 1.000 kHz
V : 1.000 V
AUTO MANU
R.H OFF
C.V OFF
INT.B OFF
OFFSET
SORT
SETTING
CALBRAT
EXIT
CAP. R/L OFFSET
SET SORT
SET PARAMETER
CALIBRATION
Press menu
key
Press F3 key
to the setting
menu
MOMORY
1
VOLT
1.000
AVGE
1
EXIT
MOMORY NO: 1
VOLTAGE = 1.000 V
AVERAGE = 1
Press F2 key to
input the desired
test voltage
Test Voltage
Figure 4-10. Programming of test voltage.
LCR-800 Series User Manual
25
4-5-6. Range Hold
If a DUT is removed from the test cable or fixture during the “Continuous” mode,
the feature of “Range Hold” can avoid range switching. Due to the test time can
be reduced, “Range Hold” is a useful utility for repetitive devices measurement.
User can just press compound key 0 to select the feature of “Range Hold”. To
disable this feature, press the same key again.
4-5-7. Averaging
If this function is enable, the testing time will multiple by the number of tests
(from 1 to 255). Hence, the accuracy can be enhanced. The measurement time
will be increased to the contrary. The programming of “Averaging” is as follows
(Figure 4-11):
l Press MENU key.
l Press F3 key to select “SETTING” menu.
l Press F3 key to select “AVGE” menu.
l Input the desired value via the numeral keys.
l Press key
l After the BAR at the bottom of LCD monitor is filled to the full, the
“Averaging” input is done.