LCR-800 Series User Manual
66
9. OPTION 2 (for LCR-816/817/819/821, RS-232 is a standard
accessory)
9-1. On-line Procedure
1) Power on the LCR METER.
2) Power on the RS232 function of LCR METER. With the steps as
follows (Figure 1):
l Press MENU key.
l Press F2 key to select SORT ” function.
l Press F3 key to select “OPTION 2” function.
l Press F1 key to select “RS232” ON.
3) Run PC LCR -VIEW Program.
4) Check the contents of the Message area under the LCR-VIEW to
make sure the Online Procedure has been well done. If not, select the
setting items above the LCR-VIEW to change the Com port and make
connection again until it is online. Please refer to Figure 9-2, 9-3, 9-4.
LCR-800 Series User Manual
67
EXIT
RS232 INTERFACE:ON
Press F1 key
to the RS232
ON
OFF
ON
SPEED
SLOW
DISPLAY
VALUE
MODE
C/D
CIRCUIT
SERIES
MENU
C 13.450uF
D .1652
TESTING
F :1.0000 kHz
V :1.000 V
AUTO MANU INT.B OFF
R.H OFF
C.V OFF
OFFSET
SORT
SETTING
CALBRAT
EXIT
CAP. C/D OFFSET
SET SORT
SET PARAMETER
CALIBRATION
Press menu
key
Press F2 key
to the sort
menu
NOM.VAL
77.000
EXIT
NOM.VAL = 77.000pF
OPTION2
Press F3 key to
input the
option2
function
RS232
Figure 9-1