6. MESSAGE CODE 
This section describes the message code for the LCR Meters.   
OVER -01 
Cause: 
1.  If the impedance of  “Device-under-test” is small than the existing 
measurement range of  the LCR Meters, the  “OVER -01 ” message will be 
displayed on the LCD monitor. 
Calculation formula:   
Capacitance: XC=1/2πfC 
Inductance: XL=2πfL 
where f=test frequency in Hz 
2.  If the inductor of  “Device-under-test ” is very large at a very high  “test 
frequency”, the  “Resonance effect ” will occur and the  impedance will 
decrease. Hence, the measured value is useless. Meanwhile, an  “OVER -01” 
message will be displayed on the LCD monitor.  
Solution: 
1.  Turn on the  “Constant Voltage” mode (please refer to  4-5-5. Constant 
Voltage Source, page 22, for details). 
2.  Select the less measurement range. Please refer to table B-1: Range Hold, 
page 42, in order to reach the exact measurement range. Turn on the “Range 
Hold” mode after the measurement range fix ed. 
Note: Both two solutions will reduce the accuracy of the LCR Meters. 
LCR-800 Series User Manual 
 
 
 
45
7. MAINTENANCE 
This section includes the basic maintenance information for the LCR Meters. 
7-1. Cleaning 
Remove the AC input power (disconnect and remove the power cord) from the 
LCR Meters before attempting to clean the instrument.   
To clear the LCR Meters, use soft cloth dampened in a solution of mild detergent 
and water. Do not spray cleaner directly onto the instrument, since it may leak 
into the cabinet and cause damage. 
Do not use chemicals containing benzine, benzene, xylene, acetone, toluene, or 
similar solvents. 
Do not use abrasive cleaners on any portion of this equipment. 
7-2. Battery Replacement 
A replaceable 3V lithium battery (BR-2/3A type) supplies the backup power of 
non-volatile memory for the LCR Meters. This battery probably needs to be 
replaced after using for 3 years.   
 
CAUTION: Danger of explosion if battery is incorrectly replaced. 
Replace only with the same or equivalent type recommended by 
the manufacturer. Discard used batteries according to the 
manufacturer’s instructions.