LCR-800 Series User Manual
42
Table A: (for range 1,2,3) –Kc
0.012≦F<0.03
35 12 9 7
0.030≦F<0.1
30 8 5 3
0.1≦F<0.25
25 6 3 2
0.25≦F<1
20 5 2 1
1 14 4 1 0
1<F≦3
15 5 2 1
3<F≦6
15 6 3 2
6<F≦10
15 8 5 3
10 <F≦20
20 10 6 5
20 <F≦50
30 22 18 15
50<F≦100
50 40 35 30
200 50 40 35 30
F: test frequency in kHz
LCR-800 Series User Manual
43
Table A: (for range 4)-Kc
0.012≦F<0.03
70 20 10 7
0.030≦F<0.1
50 13 6 3
0.1≦F<0.25
35 9 4 2
0.25≦F<1
25 6 2 1
1 15 4 1 0
1<F≦3
17 6 3 2
3<F≦6
25 15 10 6
6<F≦10
60 30 20 15
10 <F≦20
Not specified
100 65 50
20 <F≦50
50<F≦200
This range is not used above 20kHz
F: test frequency in kHz
Table B-1: Range Hold
Inductor Capacitor Resistor/Impedance
Component
Range
Max Min Max Min Max Min
Range1 16mH /f
100Ω 6.25Ω
Range2 256mH /f
1.6kΩ 0.1kΩ
Range3 4100mH /f
25.6kΩ 1.6kΩ
Range4* 65H/f 4.1H /f 6400pF /f
410kΩ 25.6kΩ
f = test frequency in kHz
*: This range is not used above 20kHz
Table B-2: Auto Range
Inductor Capacitor Resistor/Impedance
Component
Range
Max Min Max Min Max Min
Auto range
**: Above 20kHz, Cmi n = 6.4 nF/f ,and Lmax = 4100mH/f