EasyManua.ls Logo

GW Instek LCR-817 - Page 20

GW Instek LCR-817
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LCR-800 Series User Manual
34
4-5-10. Handler Interface Timing(only for LCR-826/827/829)
LCR-800 Series User Manual
35
Time Minimum
Value
Maximum
Value
T1
Trigger Pulse Width
5 us
T2
Measurement Start
Delay Time
140 us
T3
/EOM Delay Time
After Data Output
5 us
T4
Calculation and
binning Time
6 ms
SPEED
Analog Measurement Time(T5)
Trigger Wait Time
After /EOM
Output (T6)
0.012kHz
0.1kHz
0.12kHz
1kHz
10kHz100kHz
OFF
BIN
VALUE
SLOW
817ms
901ms
901ms
903ms
873ms 873ms
2ms
4ms
16ms
MEDIUM
817ms
125ms
105ms
59ms
53ms
53ms
2ms
4ms
16ms
FAST 817ms
125ms
103ms
27ms
17ms
17ms
2ms
4ms
16ms
ACCURACY
MODEL
FAST MEDIUM SLOW
LCR-827
LCR-829
LCR-826
0.5% 0.2% 0.1%
* DATA: /O_BIN_1 -_BIN_13,/O_P_OVER,/O_S_OVER,/O_S_REJ,/O_P_HI,/O_P_LO

Related product manuals