REMOTE INTERFACE 
361 
Save and Recall Commands 
 
Up to 10 different instrument states can be stored to non-volatile 
memory (memory locations 0~9).  
 
Saves the current instrument state to a specified 
save slot. When a state is saved, all the current 
instrument settings, functions and waveforms are 
also saved.  
The *SAV command doesn’t save waveforms in 
non-volatile memory, only the instrument state.  
The *RST command will not delete saved 
instrument states from memory. 
*SAV {0|1|2|3|4|5|6|7|8|9} 
Save the instrument state to memory location 0. 
 
Recall previously saved instrument states from 
memory locations 0~9. 
*RCL {0|1|2|3|4|5|6|7|8|9} 
Recall instrument state from memory location 0. 
 
Delete memory from a specified memory location. 
MEMory:STATe:DELete {0|1|2|3|4|5|6|7|8|9} 
Delete instrument state  from memory location 0.