PEL-2000 Series User Manual
142
The Pass(P)/Fail(F) Time can be set to
0.1 seconds less than the total test time.
The total test time is defined as:
On-Time + Off-Time (seconds)
If Go/NoGo is turned on but the pass
fail time is off, then Go/NoGo test will
continue, but there will not be a
specified pass/fail time window.
Short Channel: Off – 1~ 8 (CH1~CH8)
Each channel can be individually set
to simulate a short circuit (CH1~8) or
can have shorting disabled (Off).
When Short Channel is set to Off, the
channel is enabled.
7. Repeat the above steps for all ten
sequences for the same (current)
program.
Save Sequence
8. Press F3 (Save) to save all the
sequence data for the current
program.
Note
Note the program data is not yet saved into Setup
Memory. If you wish to save the Program to Setup
Memory see page 204.
Recall Default
9. To recall the Default Program
settings, press F4.
Note
If the Default is recalled, all data will be lost. This does
not include the internal Setup Memory. To see the
default settings, see page 251.