Response Speed 
 
Response speed can be set to fast or slow. Fast 
response and slow response is determined by the 
load module specifications. Slow response speeds 
are suitable for large loads as quick current 
changes will induce induction which can cause 
large voltage drops. The PEL series will try to 
rectify any voltage drops. However if voltage 
drops are too large, they may cause the load 
generator to go into oscillation. Large voltage 
drops caused by line voltage induction may 
damage the machine.   
Range Fast  Slow  
1kHz 100Hz 
Go/NoGo  
Go/NoGo testing can be with either current 
(Ampere) values (High, Low) or percentage values 
(Center, High %, Low %). A delay time of up to 1 
second can also be set.