PEL-2000 Series User Manual
86
To determine the Percent Rating (voltage or current
load input), use the following formula;

VorARating
V
VoltageExternal
InputLoad
10
Where “Rating Vor A” is the rating voltage/current
of the load module.
Current Monitor
The load current input can be externally monitored
using the IMON pin of a channel control connector.
The IMON pin outputs a voltage of 0~10V to
represent the input current of as percentage
(0~100%) of rating current.
To determine the Current Monitor Output
(IMON) , use the following formula;
V
ARating
currentinputLoad
IMON 10
Where “Rating A” is the rating current of the load
module.