SPS Power Supply User Manual
30
Rating Current Adjustment
Connection
Verification step
1. Press the AMPS key to set the
current range to high.
2. Turn the current coarse
and fine knobs fully ant-
clockwise to the minimum
positions.
&855(17
&&
&2$56(),1(
3. Turn the current coarse
and fine knobs to the
centered position
&855(17
&&
&2$56(),1(
4. Connect a multimeter as shown in the
connection diagram above.
5. Adjust VR304 so that the multimeter displays
0.00A.