PERFORMANCE ADJUSTMENT
29
Rating Voltage Adjustment
Connection
SPS Series
(rear)
Digital Multimeter
V
A
Verification step
1. Connect a multimeter as shown to the output
terminals.
2. Turn the voltage coarse
and fine knobs fully
clockwise to the
maximum positions.
3. Adjust VR301 so that the
multimeter matches the
following values
1
7
6
2
7
6
101((
1
7
6
2
7
6
101((
SPS-1820 18.50V
SPS-3610 36.50V
SPS-606 60.50V
SPS-2145 24.5V
SPS-1230 12.5V
4. Adjust VR2 so that the voltage value of the
voltage display matches the voltage shown in
the multimeter.