2.4 Leveling: The unit must sit level and solidly. Turn the leveling feet
counterclockwise to raise level. If the unit must be moved, turn the leveling
screws in all the way to prevent bending of the feet.
2.5 Cleaning: The incubator interior was cleaned at the factory, but not
sterilized. Remove all interior parts, including shelves and trays and clean
with a disinfectant that is appropriate to your application. DO NOT USE
spray cleaners that might leak through openings and cracks and get on
electrical components, or that may contain solvents that will harm coatings.
DO NOT USE chlorine-based bleaches or abrasives as this will damage
the stainless steel interior. A thorough periodic cleaning is strongly
recommended.
WARNING: Never clean the unit with alcohol or flammable cleaners with
the unit connected to the electrical supply. Always disconnect the unit from
the electrical service when cleaning and assure all volatile or flammable
cleaners are evaporated and dry before reattaching the unit to the power
supply.