SECTION 2 OPERATION
2.1 Operational Controls and Indicators
Key Description
Power key turns instrument on and off. If no keys are pressed for 5.5
minutes, the instrument turns off automatically.
Performs a measurement. To conserve battery power, the lamp turns
on only when
READ is pressed. A reading is displayed about 12
seconds after the key is pressed. After the reading is displayed, the
lamp turns off and the reading continues to be displayed. After a
measurement is taken, there is a 4-second recovery time before
another can be started. If
READ is pressed during the recovery time,
the display will flash but the lamp will not turn on until the recovery
time is complete. If no key strokes occur for 5.5 minutes, the meter
turns off.
Hold this key for continuous reading if the meter is not in the Signal
Averaging mode. After the initial delay, the reading is updated every
1.2 seconds.