3.2.3 Matching Multiple Sample Cells
Precise measurements of very low turbidity samples require the cells be
optically matched or a single cell be used for all measurements. If more
than one cell is used, follow this procedure to match (index) the cells:
1. Clean and oil
the sample cells
as instructed in
Section 3.1.1 on page 35.
2. Fill the clean sample
cells to the line with the
same sample.
3. Press: I/O to turn the
instrument on.
4. Insert the first
sample cell into the
sample compartment and
close the cover.
5. Press: READ
Record the cell position
in the compartment and
the displayed reading.
Place an orientation mark
on the cell marking band.
Note: This procedure may be
easier if the user holds the
READ key through the whole
process. This allows the lamp
to remain on and make con-
tinuous readings.
6. Insert the second
sample cell into the cell
compartment and close
the cover.
I
O
READ

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 2100P ISO and is the answer not in the manual?

Hach 2100P ISO Specifications

General IconGeneral
Instrument TypePortable Turbidimeter
Measurement Range0 to 1000 NTU
Resolution0.01 NTU on lowest range
Light SourceTungsten filament lamp
DetectorSilicon Photodiode
Operating Temperature0 to 50 °C (32 to 122 °F)
Storage Temperature-20 to 60 °C (-4 to 140 °F)
Power Supply4 AA alkaline batteries
Dimensions22.9 cm x 10.7 cm x 7.7 cm (9.0 in x 4.2 in x 3.0 in)
ComplianceISO 7027