EasyManuals Logo
Home>Hach>Measuring Instruments>2100P ISO

Hach 2100P ISO User Manual

Hach 2100P ISO
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #41 background imageLoading...
Page #41 background image
41
SECTION 3, continued
3.2.3 Matching Multiple Sample Cells
Precise measurements of very low turbidity samples require the cells be
optically matched or a single cell be used for all measurements. If more
than one cell is used, follow this procedure to match (index) the cells:
1. Clean and oil
the sample cells
as instructed in
Section 3.1.1 on page 35.
2. Fill the clean sample
cells to the line with the
same sample.
3. Press: I/O to turn the
instrument on.
4. Insert the first
sample cell into the
sample compartment and
close the cover.
5. Press: READ
Record the cell position
in the compartment and
the displayed reading.
Place an orientation mark
on the cell marking band.
Note: This procedure may be
easier if the user holds the
READ key through the whole
process. This allows the lamp
to remain on and make con-
tinuous readings.
6. Insert the second
sample cell into the cell
compartment and close
the cover.
I
O
READ

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 2100P ISO and is the answer not in the manual?

Hach 2100P ISO Specifications

General IconGeneral
BrandHach
Model2100P ISO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals