SECTION 3, continued
1. Fill the clean sample
cell to the line with
high quality water
(< 0.5 FNU). Cap and
wipe with lint-free cloth.
Apply silicone oil. See
Section 2.4.2.2 on page
25 for more information
about high quality water.
2. Press: I/O. 3. Insert the sample
cell into the sample
compartment. Close
the cover.
4. Press: READ
Record the cell position
in the compartment and
the displayed reading.
Note: This procedure may be
easier if the user holds the
READ key through the whole
process. This allows the lamp
to remain on and make
continuous readings.
5. Remove the cell,
rotate it slightly and
reinsert it into the cell
compartment. Close the
cover, then press
READ.
Record the cell position
and the displayed
reading.
6. Repeat step 5 until
the lowest reading is
displayed. Place an
orientation mark on the
cell marking band near
the top of the cell so the
cell can be consistently
inserted in the position
that yields the lowest
reading. When using the
cell, always place it in
the instrument so the
orientation mark aligns
with the raised mark on
the instrument.
I
O
READ

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 2100P ISO and is the answer not in the manual?

Hach 2100P ISO Specifications

General IconGeneral
Instrument TypePortable Turbidimeter
Measurement Range0 to 1000 NTU
Resolution0.01 NTU on lowest range
Light SourceTungsten filament lamp
DetectorSilicon Photodiode
Operating Temperature0 to 50 °C (32 to 122 °F)
Storage Temperature-20 to 60 °C (-4 to 140 °F)
Power Supply4 AA alkaline batteries
Dimensions22.9 cm x 10.7 cm x 7.7 cm (9.0 in x 4.2 in x 3.0 in)
ComplianceISO 7027